Determining Corneal Power after Radial Keratotomy.
Unlike ablative forms of myopic keratorefractive surgery (LASIK and PRK) in whichthe ratio between the posterior : anterior corneal radii is decreased, for eyes thathave previously undergone radial keratotomy, the ratio between the posterior : anteriorcorneal radii is increased. This allows for a direct estimation of the central cornealpower using elevation data of the central 4.0 mm, if carried out in a certain way.
The Zeiss Atlas 9000 MasterFit II software allows customized fitting of GP lenses according to your own fitting preferences. Zeiss Atlas Review Software is a high-value tool for existing Zeiss Atlas 993, Zeiss Atlas 995 and Zeiss Atlas 9000 users which allows you to get more out of your Zeiss Atlas Corneal Topography System. The Zeiss Atlas 995 Corneal Topographer combines comprehensive software with advanced imaging and analysis technology, making it the most popular topographer in the US. This Zeiss Atlas 995 Corneal Topographer has been fully refurbished and comes complete with a. Zeiss Atlas 995 Manual High School. 7/12/2017 0 Comments Maria Luisa Genito Apice Maria Luisa. Vittorio Emanuele, 1. Riviera del Conero. La Casa del Ghiro Pimonte. View & download of more than 429 Zeiss PDF user manuals, service manuals, operating guides. Microscope, Binoculars user manuals, operating guides & specifications. Zeiss Humphrey ATLAS 995 Corneal Topographer. We have all service manuals, instructions and intellectual property for everything we sell.
Image above: Zeiss Atlas Topographer, Annular Ring Power from the Numerical View feature. Use the average of the 1 mm, 2 mm, 3 mm and 4 mm annular power values. |
Zeiss Atlas 995 Manual 2
For eyes with prior radial keratotomy, averaging the 1 mm, 2 mm, 3 mm and 4 mm annular power rings of the Numerical View of the 995, 994, and 993 Zeiss Atlas topographer (right) will typically give a useful estimate of central corneal power.
For the Zeiss Atlas 9000 topographer, you would take average of the 1 mm, 2 mm, 3 mm and 4 mm ring (not zone) values. If the Zeiss Atlas topographer is not available, then the adjusted effective refractive power (EffRPadj) from the Holladay Diagnostic Summary of the EyeSys Corneal Analysis System can be used.
Zeiss Atlas 995 Manual Download
Above image: Axial curvature map from the Zeiss Atlas 9000 topographer displaying the ring values. For prior RK, the 1 mm, 2 mm, 3 mm and 4 mm ring values are used to estimate the central corneal power.
The key concept here is that we are looking to discover the corneal power at its center. Instruments such as manual keratometers, autokeratometers, or simulated keratometry using a standard topographer will typically over-estimate the central corneal power, resulting in a post-operative hyperopic surprise.
Of course, correctly estimating the central corneal power following RK is only half of the exercise. The calculated IOL power must also be adjusted to prevent the artifact of a very flat central corneal power from having the formula underestimate IOL power. Follow this link: 2-variable IOL Power Formulas for a summary of why this is so and how this is carried out.
Transient hyperopia following cataract surgery and prior radial keratotomy
Patients with previous 8-incision radial keratotomy will commonly show variable amounts of transient hyperopia in the immediate post-operative period following cataract surgery. This is felt to be due to stromal edema around the radial incisions, producing a temporary enhancement of central corneal flattening. While this central corneal flattening is usually transient, it can be as much as +4.00 D, and is further accentuated by greater than eight incisions, an optical zone of less than 2.0 mm, or incisions that extend all the way to the limbus.
If a patient exhibits any of the above, significant unanticipated hyperopia may be seen in the immediate post-operative period, which should gradually resolve after eight to twelve weeks. Sometimes, due to a lack of corneal stability, the post-operative refraction can continue to slowly shift myopic over a several month period. We have seen several patients with myopic shifts as large a -5.00 D over a 12-week period.
If the final post-operative refractive objective remains elusive, plans for an IOL exchange, or a piggyback IOL, should not be made until at least two months have passed and two consecutive refractions, two weeks apart (at the same time of the day), are stable (the 'rule of twos.').
Also, if more than six months passes before cataract surgery is required for the fellow eye, the corneal measurements should be repeated due to the fact that additional corneal flattening frequently occurs over time following radial keratotomy. For this reason, IOL power calculations are usually targeted for between -0.75 D and -1.00 D and are designed to make the operative eye more myopic than usual, so that five to ten years from surgery, the post-cataract surgery refractive error does not drift into hyperopia. This also helps to avoid hyperopic refractive results, which are quite common, in spite of every precaution being taken.
Of all the various forms of keratorefractive surgery, we have had the best overall accuracy following radial keratotomy using the above technique.